Precaution for handling products.
All high speed and low speed handpieces.
  • Lubricate at least twice a day and each time prior to autoclaving.
    Spray into the drive air hole until the spray emits from the
    handpiece head.

    Failure to lubricate handpiece will greatly reduce the life of the handpiece.
  • Never clean the handpiece by using oxidated potential water.
    Use of this type water could result in deterioration of metals.
  • Spray weekly with Pana Spray into whwre the bur is inserted in all push button chuck handpieces.
    This procedure will maintain the all push button chuck mechanism in good
    condition to firmly retain the bur.
  • Handpiece air pressures must be adhered to as per manufacturer´s specifications.
    Failure to comply with these specification could result in premature failure
    of the turbine.
  • Always operate with clean and dry air.
  • With low speed handpieces, before operation always confirm that the
    head is correctly and firmly attached to the shank or motor.
  • Sterilise by autoclave sterilisation.
Precaution for handling implant handpieces
  • Always clean and lubricate the handpiece shank and head after use
    and prior to sterilisation.

    Residue blood, saline solution etc. may cause deterioration to metal.
Precaution for handling laboratory handpieces
  • Use caution when using large head (4mm+) burs, wheels etc.
    Too fast speed or too greater pressure may cause the bur to break
    and cause danger to people.
  • Ensure the collet is cleaned once a week as per the instructions
    included with item when delivered.
ALWAYS READ THE INSTRUCTION MANUAL
BEFORE OPERATING ANY NEW PRODUCT.


 